福禄克校准|Fluke Calibration
7250, 7250i, and 7250xi 压力控制器/校准器 详细指标
独有的石英波登管压力传感器设计,可提供高精度的压力测量 · 提供 5 至 2500 psi 的定制压力量程 · 可实现 0.001 % FS 的稳定压力控制。
详细指标
以下指标取自福禄克校准(FLUKE CALIBRATION)官网该产品页「技术指标」表,为官网公布值;实际精度与适用条件以官方手册为准。
Technical indicators
• pressure range:Gauge pressure is optional between 40kPa~16MPa; absolute pressure is 100~400kPa and 21MPa
• Optional:Absolute pressure range using atmospheric reference sensor, from 100kPa to 17.2MPa gauge pressure, vacuum (negative gauge pressure) range from 100kPa to 17.2MPa gauge pressure
• temperature:Working temperature: 18~36℃; Storage temperature: -20~70℃
• humidity:5%~95%RH, no condensation
• Preheat time:2~3 hours; can be turned on continuously
• pressure medium:Nitrogen or clean dry air
• Test port:1/4"NPT internal thread
• Control parameters:Load capacity: 82~980cc, minimum control (absolute pressure): 1kPa absolute pressure
• communication:RS-232 and IEEE488 interface, SCPI protocol. Roska 7215 series, 7010 series, 6000 series and DPI-510 emulation mode. The LabVIEW driver can be downloaded for free from the Fluke website, and software upgrades can be completed through the RS-232 interface.
• Menu language:Chinese operation menu
• show:TFT, VGA, active matrix, 6.4 inches, 640×480 resolution, 65.000 colors
• Dimensions:All models are 18cm high x 43cm wide x 48cm deep
• weight:7.7 kg
• Atmospheric reference (optional):The maximum error is 13.79Pa per year
• Accuracy:to 17.2MPa: ±0.003%FS, 21 MPa 30% to 100% full scale 90 day: 0.012% reading 1 year: 0.013% reading
• stability:To 17.2MPa: ±0.0075%Rdg per year, 21MPa absolute pressure: ±0.01%FS per year
• control stability:Active mode: ±0.001%FS (10ppm) Passive mode: no additional uncertainty
• display resolution:User selectable, up to 1:1.000.000
• Controller response speed:When the load volume is 245cc, the controller can stabilize the output without overshoot within 15 seconds. When the 7250 range is 21MPa, the response time is 30~45 seconds.
• Negative gauge pressure accuracy:±0.003%FS
• check:The controller comes with a calibration report traceable to NIST. Maximum deviation from true pressure value including accuracy, stability, temperature effects and calibration
• total uncertainty:(~17.2MPa), 90 days: ±0.003%FS/±0.002%Rdg, 1 year: ±0.003%FS/±0.0075%Rdg Pressure range 21MPa Absolute pressure, 90 days: ±0.01% (per range) FS, 1 Year: ±0.014% (per range) FS
型号与选件
• 7250 高速数字压力控制器:
• 7250i 高速数字压力控制器:
• 7250xi 高速数字压力控制器:
资源下载
以上规格取自福禄克校准(FLUKE CALIBRATION)官网 7250, 7250i, and 7250xi 压力控制器/校准器 产品页技术指标表。