是德科技 | KEYSIGHT
M8920A 6GHz 测试组件 Upto160MHz 详细指标
Modular PXI RF Vector Transceivers
条件:document are specifications under the following conditions unless otherwise noted.
module, and 20° to 35 °C for environment temperature unless otherwise noted
Specifications describe the warranted performance of calibrated instruments. Data represented in this
• Specifications are valid from 40° to 65 °C for individual module temperature, as reported by the
• Calibrated instrument has been stored for a minimum of 2 hours within the allowed operating range
• If instrument has previously been stored at a temperature range inside the allowed storage range,
but outside the allowed operating range, instrument must have been stored for a minimum of 2 hours
within the allowed operating range before turn-on
• 30-minute warm-up time
条件:Typical describes additional product performance information that is not covered by the product warranty.
(approximately 25 °C) after alignment within the stated alignment time and temperature limits.
Nominal values indicate expected performance or describe product performance that is useful in the
application of the product but are not covered by the product warranty.
• The RF, IF, and Source Alignments have been run within the previous
7 days
• An ALL Alignment has been run within the previous
8 hours
• If the temperature has changed more than
5 °C from the previous ALL Alignment
It is performance beyond specifications that 80% of the units exhibit with a
95% confidence level. This
data, shown in italics, does not include measurement uncertainty, and is valid only at room temperature
Recommended best practices in use
• Use slot blockers and EMC filler panels in empty module slots to ensure proper operating
temperatures. Keysight chassis and slot blockers optimize module temperature performance and
• Set chassis fan to high at environmental temperatures above
45 °C.
条件:1. Instantaneous bandwidth (1 dB bandwidth) available around a center frequency over which the input signal can be digitized for
further analysis or processing in the time, frequency or modulation domain.
2. Calibration accuracy depends on how accurately the fre quency standard was adjusted to 10 MHz. If the adjustment procedure is
followed, the calibration accuracy is given by the specification. Achievable Initial Calibration Accuracy.
Option 504
100 kHz to 3.8 GHz
Option 506
100 kHz to 6 GHz
Option B40 (standard)
100 kHz – 1 MHz
1 MHz – 10 MHz
10 MHz – 20 MHz
20 MHz – 60 MHz
60 MHz – 80 MHz
80 MHz – 6 GHz
50 kHz
500 kHz
5 MHz
10 MHz
20 MHz
40 MHz
Option B10
100 kHz – 230 MHz
230 MHz – 6 GHz
Same as Option B40
80 MHz
Option B1X
100 kHz – 400 MHz
400 MHz – 6 GHz
Same as Option B10
160 MHz
RF analyzer
0.01 Hz (nominal)
Resolution bandwidth (RBW) Auto
1 Hz to 8 MHz, I/Q analyzer up to 20 MHz
Same as accuracy of internal timebase or external reference input
Internal timebase accuracy ² ± [(time since last adjustment x aging rate) ± temperature effects ± calibration
条件:Residual FM ≤ 0.25 Hz p-p in 20 ms nominal
Yearly aging rate < ± 0.1 ppm/year (± 1 x
10-7)
Achievable initial calibration accuracy ± 4 x
10-8
Temperature effects 20 to 30 °C < ± 1.5 x
10-8
Trigger Free run, external 1, external
2, RF burst, video, line, periodic
Trigger delay range
-15 to 500 ms
Amplitude range and accuracy specifications (RF T/R port and antenna port)
Antenna input port 100 kHz to 1 MHz
± 0.6 dB typical
Range: -70 dBm to +23 dBm 1 MHz to 3.9 GHz ± 1 dB
± 0.55 dB typical
3.9 GHz to 6 GHz ± 1.5 dB
± 0.95 dB typical
T/R port high power attenuation ON 100 kHz to 1 MHz
± 0.8 dB Typical
Range: -30 dBm to +47 dBm 1 MHz to 3.9 GHz ± 1 dB
± 0.85 dB Typical
3.9 GHz to 6 GHz ± 1.4 dB
± 0.95 dB Typical
T/R port high power attenuation OFF 100 kHz to 1 MHz
± 0.8 dB Typical
Range: -50 dBm to +30 dBm 1 MHz to 3.9 GHz ± 1 dB
± 0.85 dB Typical
3.9 GHz to 6 GHz ± 1.45 dB
± 0.95 dB Typical
T/R port high power attenuation ON +47 dBm (50 W) T/R port high power attenuation OFF
+33 dBm (2 W)
RF antenna port
+30 dBm (1 W)
最大值 RMS RF specified 性能电平
T/R port high power attenuation ON +47 dBm (50 W) (< 30 °C ambient), 30 W or 5 minutes on
5 minutes off
(> 30 and <
45 °C ambient)
TX post-demod audio filters
条件:Manual select filter For either LPF or HPF User Defined ³ 1. 2- pole Butterworth filter to match same filter in HP8920/8903/8901/8902. 2. 10 -pole Butterworth filter to attenuate sub- audible signaling tones. Matching the filter on HP8920/8903/8901/8902. 3. 5- pole Butterworth filter, a general -purpose filter. The 3 -dB cutoff frequency can be selected for the User -de fined audio filters.
4. T/R port high power attenuation OFF
5. Except at 100 MHz, 5 GHz, and 5.5 GHz
Low pass filter 300 Hz ¹, 3 kHz, 15 kHz, 30 kHz, 80 kHz
100 kHz, or 300 kHz
High pass filters 20 Hz, 50 Hz
300 Hz, or 400 Hz ²
Band pass filters (weighted) CCITT (ITU-T Rec.
041, ITU-T Rec. P.53), A-weighted (ANSI-IEC “A”
weighted, per IEC Rec 179), C-weighted (IEEE Std
743-1995), C-
message (C-Message per IEEE743), CCIR-1k weighted (CCIR Rec
468), CCIR-2k weighted (Dolby 2 K), CCIR unweighted (CCIR Rec
468)
Phase noise sidebands RF T/R ⁴ port and antenna port (CF =
500 MHz)
1 kHz offset
-107 dBc/Hz typical
10 kHz offset
-111 dBc/Hz typical
100 kHz offset
-108 dBc/Hz typical
1 MHz offset
-133 dBc/Hz typical
Spurious responses - Residual response in specified frequency ranges⁵
TX post-demod audio filters
Antenna input port ranged to -70 dBm
T/R port high power attenuation OFF, ranged to -70 dBm
T/R port high power attenuation ON, ranged to 0 dBm
< 200 MHz -77 dBm typical -90 dBm typical
-90 dBm typical
200 MHz to 6 GHz -95 dBm typical -90 dBm typical
-80 dBm typical
Input set to -10 dBm range <-62 dBm typical
Display Average Noise Level (DANL) with analyzer ranged to -70 dBm
T/R port high power attenuation ON 100 kHz to 1 MHz -142 dBm typical -135 dBm typical -120 dBm typical 1 MHz to 200 MHz -155 dBm, -160 dBm typical -129 dBm, -144 dBm typical -114 dBm
-127 dBm typical
200 MHz to 900 MHz -157 dBm, -161 dBm typical -144 dBm, -149 dBm typical -126 dBm
-131 dBm typical
900 MHz to 2.1 GHz -154 dBm, -158 dBm typical -140 dBm, -145 dBm typical -122 dBm
-127 dBm typical
2.1 GHz to 3.2 GHz -148 dBm, -153 dBm typical -134 dBm, -139 dBm typical -116 dBm
-121 dBm typical
3.2 GHz to 4.8 GHz -151dBm, -156 dBm typical -137dBm, -142 dBm typical -119dBm
-124 dBm typical
4.8 GHz to 6 GHz -145dBm, -150 dBm typical -129dBm, -134 dBm typical -114dBm
-119 dBm typical
Input voltage standing wave ratio (VSWR) (Attenuator =
0 dB)
RF T/R port <
1.4:1 (nominal)
Antenna port <
1.8:1 (nominal)
Trigger Free run, external 1, external
2, RF burst, video
Trigger delay range
-15 to 500 ms
Option 504
100 kHz to 3.8 GHz
Option 506
100 kHz to 6 GHz
Option B40 (standard)
100 kHz – 60 MHz
60 MHz – 220 MHz
220 MHz – 340 MHz
340 MHz – 6 Ghz
8 MHz
20 MHz
40 MHz
40 MHz
100 kHz – 230 MHz
230 MHz – 340 MHz
340 MHz – 6 GHz
Same as Option B40
40 MHz
80 MHz
Option B1X
100 kHz – 400 MHz
400 MHz – 6 GHz
Same as Option B10
160 MHz
RF Generator
0.1 Hz (nominal)
Same as accuracy of internal time base or external reference input
Accuracy¹ ± [(time since last adjustment x aging rate) ± temperature effects ±
条件:Residual FM ≤ 0.25 Hz p-p in 20 ms nominal
Yearly Aging rate < ± 0.1 ppm/year (± 1 x
10-7)
Achievable initial calibration accuracy ± 4 x
10-8
Temperature effects 20 to 30 °C < ± 1.5 ×10-8
1. Calibration accuracy depends on how accurately the frequency standard was adjusted to 10 MHz. If the adjustment procedure is followed, the calibration accuracy is given by the specification. Achievable Initial Calibration Accuracy
Amplitude level accuracy
1
Selected port levels 100 kHz to 1 MHz 1 MHz to 3.6 GHz
3.6 GHz to 6 GHz
条件:< -110 dBm ± 0.55 dB nominal ± 0.95 dB nominal ± 1.0 dB nominal
> 0 dBm ± 0.8 dB typical ± 1.5 dB, ± 0.8 dB typical ± 1.65 dB
± 1.2 dB typical
0 dBm to -60 dBm ± 0.65 dB typical ± 1.1 dB, ± 0.7 dB typical ± 1.45 dB
± 0.85 dB typical
-60 dBm to -110 dBm ± 0.65 dB typical ± 1.25 dB, ± 0.95 dB typical ± 1.6 dB
± 1.0 dB typical
条件:< -100 dBm ± 1.5 dB nominal ± 1.0 dB nominal ± 1.5 dB nominal
> -40 dBm ± 0.65 dB typical ± 1.4 dB, ± 0.8 dB typical ± 1.75 dB
± 1.4 dB typical
-40 dBm to -100 dBm ± 0.9 dB typical ± 1.4 dB, ± 0.9 dB typical ± 1.75 dB
± 1.5 dB typical
条件:< -110 dBm ± 1.5 dB nominal ± 1.0 dB nominal ± 1.5 dB nominal
1. Specifications apply when input port is set to Antenna I N.
> 0 dBm ± 0.75 dB typical ± 1.2 dB, ± 0.6 dB typical ± 1.35 dB
± 0.7 dB typical
0 dBm to -60 dBm ± 0.55 dB typical ± 1.4 dB, ± 0.75 dB typical ± 1.4 dB
± 0.85 dB typical
-60 dBm to -110 dBm ± 1.2 dB typical ± 1.3 dB, ± 1.0 dB typical ± 1.5 dB
± 0.85 dB typical
Phase noise, (CF =
500 MHz)
Gen port (+10 dBm output); T/R port (0 dBm output)
1 kHz offset <
-115 dBc/Hz typical
10 kHz offset <
-112 dBc/Hz typical
100 kHz offset <
-118 dBc/Hz typical
1 MHz offset <
-137 dBc/Hz typical
条件:T/R port high power attenuation OFF at -15 dBm 100kHz to 60MHz -34 dBc, -40 dBc nominal -35 dBc, -40 dBc nominal -40 dBc nominal 60 MHZ to 6 GHz -30 dBc, -36 dBc nominal -30 dBc, -38 dBc nominal -40 dBc nominal
+10 dBm
T/R port high power attenuation ON at 0 dBm
条件:T/R port high power attenuation OFF at -15 dBm 100kHz to 60MHz -65 dBc nominal -63 dBc nominal -65 dBc nominal 60 MHZ to 6 GHz -70 dBc nominal -70 dBc nominal -70 dBc nominal
+10 dBm
T/R port high power attenuation ON at 0 dBm
条件:> 3Ghz -40 dBc nominal -40 dBc nominal -40 dBc nominal
+10 dBm
T/R port high power attenuation ON at 0 dBm
T/R port high power attenuation OFF at -15 dBm
Input voltage standing wave ratio (VSWR)
条件:Generator port < 1.8:1 (nominal)
RF T/R port <
1.6:1 (nominal)
RF T/R port and Gen port (at –40 dBm output) 100 kHz to 60 MHz: -155 dBm typical
60 MHz to 6 GHz
< -160 dBm typical
RF T/R port 100 kHz to 6 GHz (T/R port high power attenuation OFF) –130 to +3 dBm 100 kHz to 6 GHz (T/R port high power attenuation ON)
–130 to –20 dBm, settable to –15 dBm
RF Gen port 100 kHz to 6 GHz –130 to +3 dBm 100 kHz to 6 GHz (Option M8920A-1EA)
–130 to +13 dBm, settable to +15 dBm
Frequency
0.10 Hz (nominal)
Generator maximum reverse power protection (RPP)
RF Gen port
10 W for 15 sec. with alarm
Other signaling capabilities
Signaling selection Sine, Square, Rectangle, Triangle, Dual Sine (2-tone,
M9260A PXI module Audio generator output performance Audio analyzer input performance
Channels 2 – Audio Gen Outputs
2 – Audio Analyzer Inputs
Build-in Waveforms Sine, Dual Sine Sine, Dual Sine
条件:Audio frequency accuracy¹ ± (2ppm + 100 µHz) nominal ± (2ppm + 100 µHz) nominal
Impedance 50 Ω and 600 Ω 50 Ω
600 Ω, and 1 MΩ
Audio frequency range 5 Hz to 79.8 kHz
10 Hz to 80 kHz
Audio frequency resolution 363.8 µHz
363.8 µHz
Voltage range 10 µVp to 10 Vp
10 µVp to 46Vp
Amplitude resolution 1 µV
1 µV
Channels per module 2 – BNC output channels
2 – BNC input channels
Max voltage amplitude 10 Vp output
46 Vp Input
AC accuracy ± 1.0% at 1 kHz
± 1.0% at 1 kHz
AC flatness (at 1 kHz)
5 Hz to
80 kHz BW +/- 1% (+/-0.086 dB), all ranges +/-
1% (+/-0.086 dB), all ranges
条件:Range 0 dB to 50 dB (nominal), usable up to 90 dB
Frequency range
50 Hz to 20 kHz
Voltage range
1 mV to 10 V, usable up to 46V (32 Vrms)
条件:Measurement Range 1% to 50% (nominal)
Frequency range
50 Hz to 20 kHz
Accuracy
± 0.05% (nominal)
Voltage range
1 mV to 10 V, usable up to 46V (32 Vrms)
条件:Frequency 50 Hz to 20 kHz (nominal)
Range
-90 dB to 0 dB (nominal)
Resolution
0.1 dB (nominal)
条件:Frequency 50 Hz to 20 kHz (nominal)
1. Stated frequency accuracy specification is achievable depending on the PXI chassis 10 MHz system clock accuracy or the applied
external clock reference accuracy. 2. Using two -step RMS Ratio method. Refer to M8920A Measurement Guide, Measuring RX FM Hum and Noise. For noisier signals, built -in SNR estimate method is not as accurate.
Audio module RX audio filters 1. The 3 -dB cutoff frequency can be selected for the User -defined audio filters.
Range
0 dB to 80 dB (nominal)
Low pass filter 3 kHz, 15 kHz, 30 kHz
80 kHz
High pass filters 20 Hz, 50 Hz
300 Hz, or 400 Hz
CCITT (ITU-T Rec. 041, ITU-T Rec. P.53), A-weighted (ANSI-IEC “A” weighted, per IEC Rec
179), C-
message (C-Message per IEEE743), CCIR-1k weighted (CCIR Rec 468), CCIR-2k weighted (Dolby
2 K)
五、Radio 测试 application 软件规格
N9093EM0E basic analog modulation/demodulation application key specifications
ARB generator modulation performance Analyzer demodulation performance
条件:FM deviation accuracy ± 0.2% (nominal) ± 0.1% (nominal) rate < 20 kHz
± 1.2% (nominal) rate 20 kHz to 40 kHz
Residual distortion 0.6% (nominal) ± 0.45% (nominal) rate < 20 kHz
± 1.3% (nominal) rate 20 kHz to 40 kHz
FM deviation 100 Hz to 100 kHz (settable to 1 Hz) Peak Dev
1 kHz to 100 kHz
FM Deviation Resolution 1 Hz
1 Hz
FM rate (settable range) 10 Hz to 40 kHz
10 Hz to 40 kHz
条件:AM depth accuracy ± 0.2% (nominal) depth 30% to 90%
± 2.2% (nominal) depth < 30% or >
± 0.9 (nominal) depth 10% to 95%
Residual distortion ± 0.1% (nominal) depth 30% to 90%
± 1.0% (nominal) depth < 30% or >
1.5% (nominal) depth 10% to 95%
AM depth 1% to 90%
1% to 99%
90%
AM rate (settable range) 10 Hz to 40 kHz
10 Hz to 40 kHz
90%
Single sideband modulation (USB/LSB)
条件:SSB residual distortion 1.5% (nominal) depth 10% to 95%
SSB depth 1% to 90%
1% to 99%
SSB depth Resolution 1%
1%
SSB rate range 10 Hz to 40 kHz
10 Hz to 40 kHz
SSB offset Range
300 Hz to 3 kHz
> 80 dB (SSB-SC)
条件:PM deviation accuracy ± 0.3% (nominal) 0.3 to 10 radians ± 0.3% (nominal) rate 500 Hz to 4 kHz
Residual distortion 0.4% (nominal) 0.3 to 10 radians 0.5% (nominal) dev > 1 radian
PM deviation 0.1 to 10 rad
0.2 to 10 rad
± 3.0% (nominal) rate <
500 Hz
PM resolution 0.01 rad
0.01 rad
PM rate (settable range) 10 Hz to 40 kHz
10 Hz to 4 kHz
3% (nominal) dev <
1 radian
条件:Frequency 60 Hz to 6 kHz (nominal)
Measurement range 0 dB to 50 dB (nominal)
条件:Frequency 60 Hz to 6 kHz (nominal)
Measurement range 1% to 50% (nominal)
条件:Frequency 60 Hz to 6 kHz (nominal)
Range
-90 dB to 0 dB (nominal)
Resolution
0.1 dB (nominal)
条件:Frequency 60 Hz to 6 kHz (nominal)
1. Using two -step RMS Ratio method. Refer to M8920A Measurement Guide, Measuring TX FM Hum and Noise. For noisier
Range
0 dB to 80 dB (nominal)
signals, built -in SNR estimate method is not as accurate
N9093EM1E basic digital demodulation measurement application key specifications
EVM ± 0.5% typical
± 0.6% typical
Mag error ± 0.3% typical
± 1.3% typical
FSK error ± 0.3% typical
± 1.0% typical
Deviation error ± 0.1% typical
± 1.0% typical
六、Connectivity test application software Specifications
条件:EVM at antenna input port 0.8% typical, 0.6% nominal at -10 dBm input power BW 20 MHz
EVM at antenna input port -44.5 dB typical at -20 dBm input power BW
80 MHz
-42.5 dB typical at -20 dBm input power BW
160 MHz
EVM at generator port -47 dB typical at -5 to -15 dBm output BW
80 MHz
-45 dB typical at -10 to -15 dBm output BW
160 MHz
LTE-FDD/TDD (Specifications from 700 MHz to 2.6 GHz)
ACP at antenna input port
-56 dBc typical at -10 dBm input downlink
-57 dBc typical at -10 dBm input uplink
Composite EVM at generator port 0.8% typical at -10 dBm output power BW
20 MHz
ACP at generator port
-60 dBc typical at -10 dBm output power downlink
-60 dBc typical at -10 dBm output power Uplink
条件:DEVM at generator port 1% nominal at 0 dBm output
EDR floor at antenna input port
0.6% typical at 0 dBm input power
Front panel M9470A RF interface module connections
条件:Frequency reference input 10 MHz, -5 to +10 dBm, nominal, SMB female, 50 Ω nominal
Frequency reference output 10 MHz, ± (10 MHz x frequency reference accuracy), >
0 dBm, SMB
female,50 Ω nominal
I/O characteristics M9037A controller module
USB Four USB 2.0 (type A), Two USB
3.0 ports
Ethernet Two
10/100/1000BASE-T (RJ45)
Video Dual Mode DisplayPort+ (DVI-D, VGA, HDMI with an adapter)
PCIe connector x8 PCIe connector
PXI trigger SMB (programmable direction)
Data storage (M9037A controller)
Internal
240 GB removable solid-state drive
External Supports USB
2.0 or 3.0 compatible memory devices
ARB Memory Depth information Opt. M05, 512 Msa; with a 14-bit DAC; memory requires
4 bytes per
Data Sampling information Max sampling rate is 200Msa/s, Max sampling rate for 40 MHz BW is
50
Msa/s (40*1.25).
Complies with European EMC Directive
2014/30/EU
– CISPR Pub 11 Group
1, class A
– ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
条件:altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-
Voltage/Frequency/Power (low-line) 100 to 120 V, 50/60/400 Hz, 735 W (nominal)
Voltage/Frequency/Power (high-line) 220 to 240 V, 50/60 Hz, 1300 W, (nominal)
Net 17.5 kg (38.7 lbs), nominal
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and
verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses
include, but are not limited to, temperature, humidity, shock, vibration,
Power requirements (10-slot chassis)
Weight (M9010A
10-slot chassis with basic modules)
Dimensions (M9010A
10-slot chassis)
Height
194.8 mm (with feet installed)
The M8920A Radio Test Set is supplied with a Standard one- year warranty
The recommended calibration cycle is one year; calibration services are available through Keysight service
Keysight enables innovators to push the boundaries of engineering by quickly solving
design, emulation, and test challenges to create the best product experiences. Start your
This information is subject to change without notice. © Keysight Technologies
2019 – 2023
Published in USA, October 1 , 2023 , 5992-
2802EN
Publication title Publication number
M8920A PXIe Radio Test Set Technical Overview
5992-2821EN
M8920A PXIe Radio Test Set Configuration Guide
5992-2800EN
M8920A PXIe Radio Test Set Getting Started Guide M8920-90001
M9470A PXIe 50W Interface Module Data Sheet
5992-3140EN
M9421A VXT PXIe Vector Transceiver Data Sheet
5992-1646EN
M9260A PXIe Audio Analyzer Data Sheet
5992-1918EN
KS8312xA Technical Overview for Automated Radio Testing
5992-3511EN
PXIe Chassis Spec Guide M9019-90015
PC Tested Configurations with PXIe Chassis Technical Overview
5990-7632EN
M9037A PXIe Embedded Controller Spec Guide M9037-90015
Interface Modules and Adapters for PXIe Systems
5992-0377EN
M924XA InfiniiVision PXIe Modular Oscilloscopes Data Sheet
5992-2003EN
6.5 Digit PXI Digital Multimeter Data Sheet
5992-2757EN
PXIe Vector Network Analyzer Configuration Guide
5991-4885EN
PXI Avionics Bus Interface Cards Configuration Guide
5992-2448EN
89600 VSA Software Configuration Guide
5990-6386EN
X-Series measurement applications
Signal Studio Software
89600 VSA Software:
Bluetooth and the Bluetooth logos are trademarks owned by Bluetooth SIG, Inc., U.S.A. and licensed to Keysight Technologies, Inc
以上指标摘自厂家官方数据手册,完整定义、测试条件与保证值口径以手册为准。完整指标、测试条件与定义口径请见产品页「软件和手册 | 白皮书」栏目中的官方数据手册。
是德科技 | KEYSIGHT · onetest 仪器资源库