是德科技 | KEYSIGHT
M9290A 3GHz至26.5GHz 信号分析仪 111dB 详细指标
Modular PXI Signal and Spectrum Analyzers
条件:Temperatures referred to in this document are defined as follows:
represented in this document are specifications under the following conditions unless
– Full temperature range = Individual module temperature of 5 to 68 °C, as reported by
the module, and environment temperature of 0 to 55 °C.
– Controlled temperature range = Individual module temperature of 25 to 40 °C, as
reported by the module, and environment temperature of 20 to 30 °C.
Specifications describe the warranted performance of calibrated instruments. Data
– Under auto couple control, except when Auto Sweep Time Rules = Accy
– The analyzer has been stored at an ambient temperature within the allowed
operating range for at least two hours before being turned on; if it had previously
been stored at a temperature range inside the allowed storage range, but outside
条件:enough to prevent an Alert message; if the Alert condition is changed from Time and
the uncertainties of external calibration references. These values are not warranted.
– The analyzer has been turned on at least
30 minutes with Auto Align set to normal
or, if Auto Align is set to off or partial, alignments must have been run recently
Temperature to one of the disabled duration choices, the analyzer may fail to meet
specifications without informing the user 95th percentile values indicate the breadth of the population (approx. 2σ) of perfor - mance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the
statistical observations of a sample of instruments, these values include the effects of
These values are updated occasionally if a significant change in the statistically observed
behavior of production instruments is observed
条件:units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C.
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
Typical describes additional product performance information that is not covered by
the product warranty. It is performance beyond specifications that
80 percent of the
Typical performance does not include measurement uncertainty.
Recommended best practices in use
– Use slot blockers and EMC filler panels in empty module slots to ensure proper
operating temperatures. Keysight chassis and slot blockers optimize module
– Set chassis fan to high at environmental temperatures above
45°C.
Specification Guide. The CXA-m
PXIe Signal Analyzer Specifica - tion Guide can be obtained on the web at: cxa-m_manuals
For ordering information, refer to the CXA-m PXIe Signal Analyzer
Configuration Guide (5992-0193EN).
04 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Option F03 10 Hz to 3.0 GHz
10 MHz to 3.0 GHz
Option F07 10 Hz to 7.5 GHz
10 MHz to 7.5 GHz
Option F13 10 Hz to 13.6 GHz
10 MHz to 13.6 GHz
Option F26 10 Hz to 26.5 GHz
10 MHz to 26.5 GHz
Band LO multiple (N) 0 1 10 Hz to 3.08 GHz 1 2 2.95 to 7.575 GHz 2 2
7.45 to 9.55 GHz
条件:Example frequency reference accuracy (with Option PFR)
Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Aging rate Option PFR Standard ± 1 ×10-7 / year ± 1 ×10-6 / year ± 1.5 x
10-7 / 2 years
Temperature stability Option PFR Standard 20 to 30 °C ± 1.5 ×10-8 ± 2 x
10-6
Full temperature range ± 5 ×10-8 ± 2 x
10-6
Achievable initial calibration accuracy Option PFR Standard ± 4 ×10-8 ± 1.4 x
10-6
= ± (1 x 1 ×10-7 + 5 ×10-8 + 4 x
10-8)
1 year after last adjustment = ± 1.9 x
10-7
条件:Option PFR ≤ 0.25 Hz p-p in 20 ms nominal
Standard ≤ 10 Hz p-p in 20 ms nominal
Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution
1)
Accuracy ± (marker frequency x frequency reference accuracy +
0.100 Hz)
Delta counter accuracy ± (delta frequency x frequency reference accuracy +
0.141 Hz)
Counter resolution
0.001 Hz
Range 0 Hz (zero span)
10 Hz to maximum frequency of instrument
条件:1. Horizontal resolution is span/(sweep points – 1)
Swept ± (0.25 % x span + horizontal resolution)
FFT ± (0.10 % x span + horizontal resolution)
05 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
条件:Accuracy Span ≥ 10 Hz, swept ± 0.01 % nominal
Span ≥ 10 Hz, FFT ± 40 % nominal
Range Span = 0 Hz
1 μs to 6000 s
Span ≥ 10 Hz
1 ms to 4000 s
Trigger Free run, video, external, RF burst, periodic timer
Trigger delay Span = 0 Hz or FFT
–150 to +500 ms
Span ≥ 10 Hz, swept
1 μs to 500 ms
Gate methods Gated LO; gated video; gated FFT
条件:Gate delay jitter 33.3 ns p-p nominal
Gate length range (except method = FFT)
100.0 ns to 5.0 s
Gate delay range
0 to 100.0 s
条件:Bandwidth accuracy (power) 1 Hz to 750 kHz ± 1.0 % (± 0.044 dB) nominal
820 kHz to 1.2 MHz ± 2.0 % (± 0.088 dB) nominal
1.3 to 2.0 MHz ± 0.13 dB nominal
Bandwidth accuracy (–3.01 dB) 1 Hz to 1.3 MHz ± 2 % nominal
Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10 % steps), 4, 5, 6
8 MHz
2.2 to 3 MHz
± 0.22 dB nominal
4 to 8 MHz
± 0.32 dB nominal
条件:Selectivity (–60 dB/–3 dB) 4.1:1 nominal
EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz
1 MHz (Option EMC)
EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
1 MHz (Option EMC)
Maximum bandwidth Option B25
25 MHz
条件:Local measurement and display update rate 11 ms (90/s) nominal
Remote measurement and LAN transfer rate 6 ms (167/s) nominal
Marker peak search 5 ms nominal
Center frequency tune and transfer 22 ms nominal
Measurement/mode switching 75 ms nominal
1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or
processing in the time, frequency, or modulation domain. 2. Sweep points = 101.
Range 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled
50 MHz)
Frequency and Time Specifications (Continued)
06 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Preamp off Displayed average noise level (DANL) to +30 dBm
Preamp on Displayed average noise level (DANL) to +23 dBm
Standard
0 to 70 dB in 10 dB steps
Option FSA 0 to 70 dB in 2 dB steps
7.5 GHz
0 to 70 dB in 10 dB steps
7.5 to 26.5 GHz
+30 dBm (1 W) Input attenuation ≥
10 dB, preamp off
+30 dBm (1 W) Input attenuation ≥
20 dB, preamp on
+50 dBm (100 W) < 10 μs pulse width, < 1 % duty cycle, and input attenuation ≥
30 dB
Log scale
0.1 to 1 dB/division in 0.1 dB steps
1 to 20 dB/division in 1 dB steps (10 display divisions)
Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A
条件:(10 dB input attenuation, 20 to 30 °C, σ = nominal standard deviation)
(reference setting) 100 kHz to 3.0 GHz ± 0.30 dB nominal 3.0 to 7.5 GHz ± 0.50 dB nominal 7.5 to 26.5 GHz ± 0.70 dB nominal
Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, σ = nominal standard deviation)
Frequency response Specification 95th percentile (≈
2 σ)
9 kHz to 10 MHz ± 0.50 dB
± 0.4 dB
10 MHz to 3 GHz ± 0.65 dB
± 0.5 dB
3 to 13.6 GHz ± 1.30 dB
± 0.8 dB
13.6 to 19.3 GHz ± 1.50 dB
± 1.0 dB
19.3 to 24.2 GHz ± 2.20 dB
± 1.3 dB
24.2 to 26.5 GHz ± 2.50 dB
± 1.3 dB
Preamp on (Option P03, P07, P13, P26)
100 kHz to 10 MHz
± 0.5 dB
Input attenuation switching uncertainty Specifications Additional information
Attenuation > 2 dB, preamp off 50 MHz (reference frequency) ± 0.3 dB
± 0.15 dB typical
Relative to 10 dB
07 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
At all frequencies ± (0.40 dB + frequency response)
100 kHz to 3 GHz ± 0.60 dB (95th Percentile ≈
2σ)
Preamp on (Option P03/P07/P13/P26) ± (0.36 dB + frequency response) (95th percentile)
Preamp off (10 dB attenuation) Specification Additional information
9 kHz to 10 MHz ± 0.70 dB
± 0.31 dB typical
10 MHz to 3 GHz ± 0.85 dB
± 0.40 dB typical
3 to 7.5 GHz ± 0.95 dB
± 0.44 dB typical
7.5 to 9.55 GHz ± 0.90 dB
± 0.38 dB typical
9.55 to 12.55 GHz ± 0.95 dB
± 0.44 dB typical
12.55 to 13.6 GHz ± 0.90 dB
± 0.32 dB typical
13.6 to 15.55 GHz ± 0.95 dB
± 0.35 dB typical
15.55 to 22.8 GHz ± 1.10 dB
± 0.50 dB typical
22.8 to 24.2 GHz ± 1.20 dB
± 0.50 dB typical
24.2 to 26.5 GHz ± 1.30 dB
± 0.57 dB typical
Preamp on (0 dB attenuation)
100 kHz to 10 MHz
± 0.30 dB typical
10 MHz to 3 GHz
± 0.65 dB typical
3 to 7.5 GHz
± 0.73 dB typical
7.5 to 9.55 GHz
± 0.49 dB typical
9.55 to 12.55 GHz
± 0.71 dB typical
12.55 to 19.3 GHz
± 0.65 dB typical
19.3 to 21 GHz
± 0.80 dB typical
21 to 22.8 GHz
± 1.20 dB typical
22.8 to 24.2 GHz
± 1.39 dB typical
24.2 to 26.5 GHz
± 1.66 dB typical
Input voltage standing wave ratio (VSWR) (≥
10 dB attenuation)
10 MHz to 3 GHz <
1.2 nominal
3 to 7.5 GHz <
1.4 nominal
7.5 to 13.6 GHz <
1.6 nominal
13.6 to 24.2 GHz <
1.8 nominal
24.2 to 26.5 GHz <
2.2 nominal
Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW)
1 Hz to 3 MHz RBW
± 0.15 dB
4, 5, 6, 8 MHz RBW
± 1.0 dB
Log scale
–170 to +23 dBm in 0.01 dB steps
Linear scale Same as log (707 pV to 3.16 V)
Switching between linear and log
0 dB
Log scale/div switching
0 dB
Amplitude Accuracy and Range Specifications (Continued)
08 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
–80 dBm ≤ input mixer level < –10 dBm
± 0.15 dB total
条件:Gain 100 kHz to 26.5 GHz +20 dB nominal
Noise figure 10 MHz to 3 GHz 10 dB nominal
3 to 26.5 GHz DANL + 176.24 dB nominal
Normal, peak, sample, negative peak, log power average, RMS average, and voltage average
Preamplifier (Option P03/P07/P13/P26)
Frequency range Option P03
100 kHz to 3.0 GHz
Option P07
100 kHz to 7.5 GHz
Option P13
100 kHz to 13.6 GHz
Option P26
100 kHz to 26.5 GHz
Amplitude Accuracy and Range Specifications (Continued) 09 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
1 dB gain compression (two-tone)
条件:Preamp off 10 MHz to 7.5 GHz +6 dBm nominal
Preamp on 10 MHz to 7.5 GHz –15 dBm nominal
7.5 to 26.5 GHz –19 dBm nominal
7.5 to 26.5 GHz
+4 dBm nominal
Displayed average noise level (DANL) (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High
20 to 30 °C)
Parentheses indicate typical performance
条件:Preamplifier OFF Preamplifier ON 10 Hz –95 dBm nominal 100 Hz –110 dBm nominal 1 kHz –115 dBm nominal
9 kHz to 1 MHz (–125) dBm
1 to 10 MHz –144, (–148) dBm
–154, (–158) dBm
10 MHz to 1.5 GHz –148, (–150) dBm
–160, (–163) dBm
1.5 to 4.5 GHz –146, (–149) dBm
–160, (–163) dBm
4.5 to 7 GHz –141, (–145) dBm
–157, (–161) dBm
7 to 9.5 GHz –144, (–147) dBm
–158, (–160) dBm
9.5 to 13 GHz –136, (–140) dBm
–156, (–160) dBm
13 to 14.5 GHz –142, (–145) dBm
–158, (–161) dBm
14.5 to 19.3 GHz –132, (–138) dBm
–153, (–157) dBm
19.3 to 23 GHz –134, (–139) dBm
–152, (–157) dBm
23 to 24 GHz –132, (–137) dBm
–150, (–155) dBm
24 to 26.5 GHz –128, (–133) dBm
–144, (–149) dBm
Residual response (Input terminated and 0 dB attenuation) 200 kHz to 26.5 GHz (swept)
–90 dBm
Zero span or FFT or other frequencies
–100 dBm nominal
Tuned frequency (f) Excitation Freq Mixer level Response
Image responses (Second mixer) 10 MHz to 7.5 GHz f + 1645 MHz −10 dBm −70 dBc (−80 dBc typical) 7.5 GHz to 19.3 GHz f - 1645 MHz −10 dBm −70 dBc (−80 dBc typical) 19.3 GHz to 21 GHz f + 1645 MHz −10 dBm
−70 dBc (−80 dBc typical)
21 GHz to 22.8 GHz f - 1645 MHz −10 dBm
−70 dBc (−80 dBc typical)
22.8 GHz to 24.2 GHz f + 1645 MHz −10 dBm
−70 dBc (−80 dBc typical)
24.2 GHz to 26.5 GHz f - 1645 MHz −10 dBm
−70 dBc (−80 dBc typical)
LO-related spurious 10 MHz to 26.5 GHz −10 dBm
−64 dBc typical
Other spurious responses Mixer level Response
条件:First RF order (f ≥ 10 MHz from carrier) −10 dBm −70 dBc (−80 dBc nominal)
High RF order (f ≥ 10 MHz from carrier) −30 dBm −70 dBc (−80 dBc nominal)
IF feedthrough −10 dBm
−75 dBc (−80 dBc typical)
10 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Parentheses indicate typical performance
条件:10 MHz to 26.5 GHz –8 dBm nominal
Figure 1. Nominal dynamic range for Band 0,
Figure 2. Nominal dynamic range, for
Preamp off (Two –20 dBm tones at input mixer spaced by 100 kHz, 0 dB attenuation
20 to 30 °C
10 MHz to 2 GHz
+12 dBm, (+16) dBm
2 to 3 GHz
+12 dBm, (+17) dBm
3 to 7.5 GHz
+12 dBm, (+16) dBm
7.5 to 13.6 GHz
+11 dBm, (+15) dBm
13.6 to 26.5 GHz
+10 dBm, (+14) dBm
Option P03/P07/P13/P26 Preamp on
(Two
–45 dBm tones at the preamp input
spaced by 100 kHz
0 dB attenuation
20 to 30 °C)
for second and third order distortion
10 MHz to 3 GHz
second and third order distortion
3 GHz to
26.5 GHz
Dynamic Range Specifications (Continued)
11 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Noise sidebands (20 to 30 °C, CF =
1 GHz)
1 kHz –102 dBc/Hz
–105 dBc/Hz
10 kHz –106 dBc/Hz
–110 dBc/Hz
100 kHz –108 dBc/Hz
–110 dBc/Hz
1 MHz –130 dBc/Hz
–132 dBc/Hz
10 MHz
–145 dBc/Hz nominal
条件:Figure 3. Nominal phase noise at different
center frequencies
12 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Amplitude accuracy, W-CDMA or IS95
(20 to 30 °C, attenuation =
10 dB)
± 1.23 dB (± 0.62 dB
95th percentile)
Accuracy, W-CDMA (ACLR) (at specific mixer levels and ACLR ranges)
Without noise correction –68 dB
–72 dB
With noise correction –73 dB
–78 dB
Offset channel pairs measured
1 to 6
Multiple number of carriers measured Up to 12
Histogram resolution
0.01 dB
Maximum harmonic number
10th
Results Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in %
Measure the third-order products and intercepts from two tones
Methods Power above threshold, power within burst width
Results Single burst output power, average output power, maximum power, minimum power within burst, burst
W-CDMA (1 to 3.0 GHz) table-driven spurious signals; search across regions
Dynamic range
75.0 dB (79.1 dB typical)
Absolute sensitivity
−82.5 dBm (−86.5 dBm typical)
Spectrum emission mask (SEM) cdma2000 ® (750 kHz offset)
Relative dynamic range (30 kHz RBW)
73.0 dB (78.1 dB typical)
Absolute sensitivity
−99.7 dBm (−102.7 dBm typical)
Relative accuracy
± 0.11 dB
3GPP W-CDMA (2.515 MHz offset)
Relative dynamic range (30 kHz RBW)
77.5 dB (82.4 dB typical)
Absolute sensitivity
−97.7 dBm (−101.7 dBm typical)
Relative accuracy
± 0.15 dB
13 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Option T07
2 MHz to 7.5 GHz
Option T13
2 MHz to 13.6 GHz
Option T26
2 MHz to 26.5 GHz
条件:Absolute accuracy (at 50 MHz, –15 dBm) ± 1.0 dB typical ± 0.3 dB nominal
Level accuracy Nominal 2 MHz to 7.5 GHz ± 0.8 dB 7.5 to 13.6 GHz ± 0.9 dB 13.6 to 23GHz ± 1.5 dB
Range 2 MHz to 10 GHz —35 to 0 dBm 10 to 20 GHz —35 to —5 dBm 20 to 26.5 GHz
—40 to —12 dBm
Output flatness (Referenced to 50 MHz, –15 dBm) 95th percentile (≈
2σ)
2 MHz to 7.5 GHz ± 1.0 dB
± 0.3 dB
7.5 to 13.6 GHz ± 1.2 dB
± 0.3 dB
13.6 to 23 GHz ± 1.8 dB
± 0.6 dB
23 to 26 GHz ± 2.5 dB
± 1.2 dB
Range 2 MHz to 10 GHz —35 to 0 dBm 10 to 20 GHz —35 to —5 dBm 20 to 26.5 GHz
—40 to —12 dBm
Average total power +
30 dBm (1 W)
Noise sidebands (Center Frequency =
1 GHz) Offset Typical
Complies with European EMC Directive
2004/108/EC
– CISPR Pub 11 Group
1, class A
– ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
条件:and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environ -
mental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude,
Power drawn from chassis ≤
65 W
The recommended calibration cycle is one year; calibration services are available through Keysight service centers
条件:Connector 3.5mm female, 50 Ω nominal
Connector SMB male, 50 Ω nominal
Connector SMB male, 50 Ω nominal
10 MHz in
10 MHz out
条件:Connector SMB male, 50 Ω nominal
Noise source For use with Keysight
346 and SNS series noise sources
15 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
Span = 1 MHz
50 Hz to 1 MHz
Span = 10 kHz
1 Hz to 10 kHz
Span = 100 Hz
100 MHz to 100 Hz
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B
110 dB)
条件:Center frequency (GHz) Span (MHz) Max. error RMS (nominal)
IF phase linearity (deviation from mean phase linearity, nominal)
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
IF frequency response (standard
10 MHz IF path)
IF frequency response (demodulation and FFT response relative to the center frequency
20 to 30 °C)
≤ 3.0 ≤ 10 ± 0.40 dB
0.03 dB
3.0 < f ≤ 26.5 ≤ 10
0.10 dB
中心频率 (GHz) 跨度 (MHz) 峰值-to-peak RMS
条件:Center frequency (GHz) Span (MHz) Max. error RMS (nominal)
IF phase linearity (deviation from mean phase linearity, nominal)
3.0 < f ≤ 7.5 ≤ 10 0.5 °
0.4 °
Data acquisition (standard
10 MHz IF path)
Time record length
4,000,000 IQ sample pairs
Option B25
25 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency
20 to 30 °C)
≤ 3.0 10 to ≤ 25 ± 0.45 dB
0.03 dB
3.0 < f ≤ 26.5 10 to ≤ 25
0.20 dB
Center frequency (GHz) Span (MHz) Peak-to-peak RMS 0.02 ≤ f < 3.0 10 to ≤ 25 1 ° 0.3 ° 3.0 < f ≤ 7.5 10 to ≤ 25 1 °
0.5 °
Data acquisition (B25 IF path)
IQ analyzer
4,000,000 IQ sample pairs
Operating system Windows 7 (32 &
64 bit)
Processor speed
1.86 GHz minimum
2.4 GHz recommended
Available memory
4 GB minimum
8 GB recommended
Video Support for DirectX
9 graphics with 128 MB graphics recommended (SuperVGA supported)
Browser Microsoft Internet Explorer
7.0 or greater
16 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
Instrument connection software
Keysight IO library The IO library suite offers a single entry point for connection to
the most common instruments including AXIe, PXI, GPIB, USB,
Ethernet/LAN, RS-232, and VXI test instruments from Keysight and
other vendors. It automatically discovers interfaces, chassis, and
instruments. The graphical user interface allows you to search for,
verify, and update IVI instrument and soft front panel drivers for
modular and traditional instruments. The IO suite safely installs in
The CXA-m includes a soft front panel (SFP), a software-based
graphical user interface (GUI) which enables the instrument’s
Keysight connection expert Connection expert is the graphical user interface included in the
IO libraries suite that allows you to search for, verify and update IVI
instrument and soft front panel drivers for modular and traditional
Driver Development environments lVI-COM Visual Studio (VB .NET, C#, C/C++), VEE, LabVIEW, LabWindows/
条件:syntax checking, command execution, and debug tools to build sequences for integration in Excel, MATLAB, Visual Studio, and VEE.
Command expert Assists in finding the right instrument commands and setting correct
parameters. A simple interface includes documentation, examples,
X-Series measurement applications
条件:Provides measurements for analog demodulation, noise figure, phase noise, vector signal analysis, and others.
Licensed software. For more information, visit find/x-series_apps 89600 VSA 89600 VSA software sees through the complexity of emerging and
existing industry standards, serving as your window into complex
Licensed software. For more information, visit
SystemVue SystemVue is a system-level EDA platform for designing communi -
cations and defense systems. Used with the M9290A, SystemVue
enables you to create model-based design validation tests to ensure
Licensed software. For more information, visit systemvue
Software
17 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
M9290A CXA-m PXIe Signal Analyzer - Product Fact Sheet
5992-0044EN
M9290A CXA-m PXIe Signal Analyzer - Configuration Guide
5992-0193EN
M9018A PXIe 18 slot Chassis - Data Sheet
5990-6583EN
M9037A PXIe High Performance Embedded Controller - Data Sheet
5991-3661EN
M9036A PXIe Embedded Controller - Data Sheet
5990-8465EN
For more information or literature resources please visit the web:
以上指标摘自厂家官方数据手册,完整定义、测试条件与保证值口径以手册为准。完整指标、测试条件与定义口径请见产品页「软件和手册 | 白皮书」栏目中的官方数据手册。