是德科技 | KEYSIGHT
EDUX1052G 示波器 50MHz 详细指标
Benchtop Oscilloscopes
Input coupling DC, AC (10 Hz cutoff frequency)
Input impedance/capacitance 1 MΩ ± 2%, 16 pF
±3 pF
Input sensitivity range4
500 μV/div to 10 V/div
Standard probes N2142A 1/10 switchable
75 MHz (2 included in EDUX1052A/EDUX1052G)
N2140A 1/10 switchable
200 MHz (2 included in DSOX1202A/DSOX1202G)
N2140A 1/10 switchable
200 MHz (4 included in DSOX1204A/DSOX1204G)
Probe attenuation factor
0.1X to 10,000X in 1-2-5 sequence; (–20 dB to +80 dB in 0.1 dB steps)
Hardware bandwidth limits Approximately
20 MHz (selectable)
Vertical resolution
8 bits
条件:1. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and from ± 10 °C user calibration temperature.
2. Bandwidth specifications apply for 1 mV/div to 10 V/div vertical settings. Bandwidth at the 500 μV/div vertical setting is limited to 20 MHz.
3. Half-channel operation on a 2-channel model refers to two-channel operation when using channel-1 or channel-2 AND channel-3 or channel-4.
4. 500 μV/div is a 2X digital magnification of 1 mV/div setting.
Maximum input voltage 150 Vrms
200 Vpk
DC vertical accuracy ± [DC vertical gain accuracy + DC vertical offset accuracy +
0.25% full scale]
DC vertical gain accuracy 1 +3% full scale (≥
10 mV/div)
+4% full scale (<
10 mV/div)
DC vertical offset accuracy ± 0.1 div ± 2 mV
± 1% of offset setting
Skew Channel to channel
1 ns (without deskew)
Channel to external
2 ns (without deskew)
Offset range 500 uV/div to 200 mV/div
+2 V
> 200 mV/div to 10 V/div
+100 V
适用型号 EDUX1052A/EDUX1052G DSOX1202A/DSOX1202G DSOX1204A/DSOX1204G
Bandwidth (–3 dB) 1, 2 50 MHz 70 MHz
70 MHz
100 MHz (option D1202BW1A)
100 MHz (option D1200BW1A)
200 MHz (option D1202BW2A)
200 MHz (option D1200BW2A)
Calculated rise time (10 to 90%) ≤ 7 ns ≤ 5 ns (70 MHz base model) ≤
5 ns (70 MHz base model)
≤ 3.5 ns (with 100 MHz option) ≤ 3.5 ns (with
100 MHz option)
≤ 1.7 ns (with 200 MHz option) ≤ 1.7 ns (with
200 MHz option)
Maximum sample rate 1 GSa/s (all channels)
2 GSa/s (all channels)
1 GSa/s (if ext. trigger is displayed)
2 GSa/s (one- or half-channel3 operation)
1 GSa/s (three- or four-channel operation)
Maximum memory depth 200 k points (all channels)
2 M points (all channels)
1 M points (if ext. trigger is displayed)
2 M points (one- or half-channel3 operation)
1 M points (three- or four-channel operation)
Waveform update rate ≥ 100,000 waveforms/sec ≥ 200,000 waveforms/sec ≥
200,000 waveforms/sec
Time base range 5 ns/div to 50 s/div (50 MHz, 70 MHz , and 100 MHz models)
2 ns/div to 50 s/div (200 MHz models)
Horizontal resolution
2.5 ps
Timebase accuracy 5 50 ppm
± 5 ppm per year (aging)
Timebase delay time range Pre-trigger: Greater of 1 screen width or 200 μs
Channel to channel deskew range
± 100 ns
Δ Time accuracy (using cursors) ± (time base acc. x reading) ± (0.0016 x screen width)
± 200 ps (same channel)
Modes Main, zoom, roll, XY
条件:Acquisition system 5. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and from ± 10 °C user calibration temperature.
XY X = channel 1, Y = channel 2, Z = external trigger
1.4 V blanking
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: <
0.5 degree
Maximum sample rate 1 GSa/s 2 GSa/s (2 ch operation)
1 GSa/s (4 ch operation)
Maximum record length 200 k points 2 M points (2 ch operation)
1 M points (4 ch operation)
Acquisition 模式正常 Default 模式 Default 模式
Peak Detect Capture glitches as narrow as
10 ns at all time base
Capture glitches as narrow as: 70 MHz model: 10 ns at all time base settings 100 MHz model: 5 ns at all time base settings 200 MHz model
2.5 ns at all time base setting
Averaging Selectable from 2, 4, 8, 16, 64, ... to 65,536 Selectable from 2, 4, 8, 16
64, ... to 65,536
High Resolution Real-time boxcar averaging reduces random noise
and effectively increases vertical resolution to 12 bits
of resolution when ≥
20 μs/div at 1 GSa/s
Real-time boxcar averaging reduces random noise and effectively increases vertical resolution to 12 bits of resolution when ≥
20 μs/div at 2 GSa/s
Segmented Not available Segmented memory optimizes available memory for
data streams that have long dead times between activity
Maximum number of segments =
500
Minimum trigger re-arm time =
1 µs (1,000,000
时间模式正常 Default 模式 Default 模式
条件:Optimizes vertical scaling for stacked waveforms and sets timebase to display ~ 1.8 periods. Can be customized to function on just channels that are previously turned on and displayed.
Optimizes vertical scaling for stacked waveforms and sets timebase to display ~ 1.8 periods. Can be customized to function on just channels that are previously turned on and displayed.
Roll Displays the waveform moving across the screen from right to left. Available at the timebase settings of 50 ms/div or slower
Displays the waveform moving across the screen from right to left. Available at the timebase settings of 50 ms/div or slower
XY Displays the volts-versus-volts display Displays the volts-versus-volts display
X = Channel 1, Y = Channel 2 X = Channel 1, Y = Channel
2
Z = External trigger, 1.4 V blanking Z = External trigger
1.4 V blanking
Phase error at 1 MHz: < 0.5 degree Phase error at 1 MHz: <
0.5 degree
Autoscale Finds and displays all signals connected to analog input channels and the external trigger input. Sets trigger type to rising edge at ~50% on external (highest priority source), or lowest numbered
channel with a signal that exceeds
~10 mVpp.
Finds and displays all signals connected to analog input channels and the external trigger input. Sets trigger type to rising edge at ~50% on external (highest priority source), or lowest numbered
channel with a signal that exceeds
~10 mVpp.
Trigger sources Analog channels, line6, external, WaveGen, WaveGen modulation FM/FSK
Trigger modes Normal (triggered): Requires trigger event for oscilloscope to trigger
Auto: Triggers on selected source or automatically triggers (asynchronously) in absence of a valid trigger event
Single: Triggers only once upon detection of a valid trigger event
Force: Front panel button that forces an asynchronous trigger while in the Normal trigger mode
AC: AC coupled trigger, cutoff frequency: ~
10 Hz
HF reject: High frequency reject, cutoff frequency ~
50 kHz
LF reject: Low frequency reject, cutoff frequency ~
50 kHz
Noise reject: Selectable OFF or ON, decreases trigger sensitivity
2X
Trigger holdoff range
60 ns to 10 s
MHz)
0.9 div or 3.8 mV (10 to 50
Greater of: 0.6 div or 2.5 mV (≤ 10 MHz) 0.9 div or 3.8 mV (10 to 70 MHz)
1.2 div or 5 mV (70 to 200 MHz)
External ≤ 10 MHz: 250 mVpp ≤
10 MHz:
20 mVpp (1.6 V range)
100 mVpp (8 V range)
10 to 50 MHz: 500 mVpp
10 to 200 MHz:
100 mVpp (1.6 V range)
500 mVpp (8 V range)
Internal ± 6 div from center-screen
± 6 div from center-screen
External 8 ± 8 V ± 1.6 V or
± 8 V selectable
Edge Trigger on a rising, falling, alternating or either edge of any source
Pattern/state Not available Trigger when a specified pattern/state on any
combination inputs is entered
9
Pulse width Trigger on a pulse of a selected channel with a time duration that is ‘less than a value,’
‘greater than 值’ 或 ‘inside 时间范围’
Range minimum: 10 ns
10 s max
Setup and hold Not available Trigger and clock/data setup and/or hold time
violation. Setup time can be set from –7 ns to 10 s.
Hold time can be set from 0 s to 10 ns
Rise/fall time Not available Trigger on rise-time or fall-time edge-speed violations
(< or >) based on a user-selectable threshold and time setting range between
5 ns and 10 s
Video Trigger on all lines or individual lines; odd/even or all fields from the composite video; or
broadcast standards (NTSC, PAL, SECAM, and PAM-M)
条件:I2C Trigger at a start/stop condition or user-defined frame with address and/or data values. Also,
trigger on missing acknowledge, restart, EEPROM read and 10-bit write
RS-232/422/485/UART
Trigger on Rx or Tx start bit, stop bit, data content or parity error
Not available Trigger on SPI (Serial Peripheral Interface) data
pattern during a specific framing period
Supports positive and negative chip select framing as well as clock idle framing. Supports
MOSI or MISO (4-channel models) data as half
Not available Trigger on CAN (controller area network) version
2.0A and 2.0B signals. Trigger on the start of
Not available Trigger on LIN (Local Interconnect Network) sync
break, frame ID, frame ID + data, parity error, or checksum error
I2C Baud Rate: Up to 3.4 Mbps
Address size
7-bit or 8-bit
One plus protocol lister/table
Baud rate: Up to 3.4 Mbps
Address size
7-bit or 8-bit
One plus protocol lister/table
UART/RS232 Baud Rate
100 bps to 10 Mbps
Two (Tx and Rx) plus protocol lister/table
Baud rate
100 bps to 10 Mbps
Two (Tx and Rx) plus protocol lister/table
SPI10 Not available Baud rate: Up to 25 Mbps
Chip select: 低, 高, 或时间-out
Number of time-correlated decode traces on
4-channel models:
Two (MISO and MOSI) plus protocol lister/table
Number of time-correlated decode traces on
2-channel models:
One (Data) plus protocol lister/table
CAN Not available Baud rate
10 kbps to 5 Mbps
Standard: “Classic” CAN
2.0
Real-time totalizer: Number of frames, number of error frames,
number of overload frames, bus load (%)
条件:10. 4-channels models (DSOX1204A or DSOX1204G) recommended for 4-wire SPI measurement applications.
Number of time-correlated decode traces: One plus protocol lister/table
LIN Not available Baud rate
2.4 kbps to 625 kbps
Standards: LIN
1.3 and 2.x
Number of time-correlated decode traces: One plus protocol lister/table
Cursors Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy +
0.25% full scale]
Dual cursor accuracy: ± [DC vertical gain accuracy +
0.5% full scale]
Units: Seconds(s), Hz (1/s), phase (degrees)
Automatic measurements Select up to 4 continuously updated measurements from a list of 32 available amplitude, timing, and count measurements
Cursors track last selected measurement
Use default (relative/%) or customizable measurement threshold levels (absolute or relative)
Measurements automatically gated by zoom window
Vertical/amplitude measurements (14):
Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, preshoot, average-N cycles, average-full screen,
DC RMS-N cycles, DC RMS-full screen, AC RMS-N cycles, AC RMS-full screen (standard deviation)
Timing measurements (14):
Period, frequency, counter, +width, -width, +duty cycle, -duty cycle, bit rate, rise time, fall time,
延迟, 相位, X · 最小 Y, X · 最大 Y
Count measurements (4):
+pulse count, -pulse count, rising edge count, falling edge count
Performs
24 parametric measurements once (not updated) on a single source (ch1, ch2, ch3, or ch4) one time
Automatic measurement logging: Available via BenchVue BV0004B (standard)
Math functions Add, subtract, multiply, divide, FFT (magnitude), FFT (phase), low-pass filter
Record size Up to 64 k points resolution
FFT Window types: Hanning, Flat top, Rectangular, Blackman-Harris
Vertical scaling: dB (logarithmic) or RMS (linear)
Horizontal scaling: User-defined span and center frequency settings, or Auto Setup
Functions DC, AC-rms, DC-rms
Measuring rate
100 times/second
Auto ranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter Graphical display of most recent measurement, plus extrema over the previous
3 seconds
Measuring rate
100 times/second
Auto ranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter Graphical display of most recent measurement, plus extrema over the previous
3 seconds
Frequency response analysis - Bode plot (standard on “G” models)
适用型号 EDUX1052G/ DSOX1202G/ DSOX1204G
Dynamic range > 80 dB (typical, based on 0 dBm (630 mVpp) input into
50-Ω load
VIN and VOUT Channel 1, 2, 3, and 4 (channel 3 and 4 on
4-channel models only)
Frequency range
10 Hz to 20 MHz
Number of test points
1 to 1000 points across selected frequency range
Test amplitude 1 mVpp to 9 Vpp into
50-Ω
Test results Overlaid logarithmic gain (dB) and linear phase (degrees) plots versus logarithmic frequency
Manual measurements A single pair of tracking gain and phase markers at user-defined frequency setting
Plot scaling Auto-scaled during test with user-defined scaling after test
条件:Note: Only available on WaveGen models EDUX1052G, DSOX1202G, and DSOX1204G. WaveGen is not upgradeable.
WaveGen – Built-in function generator (standard on “G” models)
Carrier waveforms: Sine, ramp
Modulation source: Internal (no external modulation capability)
– Modulation: sine, square, ramp
– Modulation frequency
1 Hz to 20 kHz
– Modulation: sine, square, ramp
– Modulation frequency
1 Hz to 20 kHz
– Minimum carrier frequency
10 Hz
– Deviation
1 Hz to carrier frequency or (2e12 / carrier frequency), whichever is smaller
– Modulation
50% duty cycle square wave
– Hop frequency
2 x FSK rate to 10 MHz
Sine Frequency range
0.1 Hz to 20 MHz
Amplitude flatness
± 0.5 dB (relative to 1 kHz)
Harmonic distortion
—40 dBc
Spurious (non-harmonics)
—40 dBc
Total harmonic distortion
1%
SNR (50 Ω load, 500 MHz bandwidth): 40 dB (typical)
30 dB (min)
Square wave /pulse Frequency range
0.1 Hz to 10 MHz
Duty cycle resolution: Larger of 1% or 10 ns
Rise/fall time
18 ns (10 to 90%)
Pulse width resolution
10 ns or 5 digits, whichever is larger
Asymmetry (at 50% DC): ± 1%
± 5 ns
Ramp /triangle wave Frequency range
0.1 Hz to 200 kHz
Variable symmetry
0 to 100%
Noise Bandwidth
20 MHz typical
适用型号 EDUX1052G/ DSOX1202G/ DSOX1204G
Frequency Sine wave and ramp accuracy: 130 ppm (frequency < 10 kHz) 50 ppm (frequency >
10 kHz)
[50 + frequency/200] ppm (frequency <
25 kHz)
50 ppm (frequency ≥
25 kHz)
Resolution
0.1 Hz or 4 digits, whichever is larger
Amplitude Square, Pulse, Ramp: 2 mVpp to 20 Vpp into Hi-Z (offset ≤ ±0.4 V) 1 mVpp to 10 Vpp into 50 Ω (offset ≤ ±0.4 V) 50 mVpp to 20 Vpp into Hi-Z (offset >
±0.4 V)
25 mVpp to 10 Vpp into 50 Ω (offset >
±0.4 V)
Sine: 2 mVpp to 12 Vpp into Hi-Z (offset ≤ ± 0.4 V) 1 mVpp to 9 Vpp into 50 Ω (offset ≤ ± 0.4 V) 50 mVpp to 12 Vpp into Hi-Z (offset >
± 0.4 V)
25 mVpp to 9 Vpp into 50 Ω (offset >
± 0.4 V)
Resolution: ≤
1% of the amplitude
Accuracy: 2% (Frequency =
1 kHz)
DC offset Square, Pulse, Ramp: ± [10 V – ½ amplitude] into Hi-Z ± [5 V – ½ amplitude] into
50 Ω
Sine: ± [8 V – ½ amplitude] into Hi-Z ± [4.5 V – ½ amplitude] into
50 Ω
Resolution: Larger of 100 µV or 3 digits
Accuracy: ± 1.5% of offset setting ± 1.5% of amplitude
± 1 mV
Main output Impedance
50 Ω typical
Standard Ports One USB
2.0 hi-speed device port on rear panel. Supports USBTMC protocol
One USB
2.0 hi-speed host port on front panel. Supports memory devices
One Ethernet
10/100Base-T port on rear panel
条件:Maximum Relative Humidity (non-condensing): 95%RH up to 40°C, decreases linearly to 45%RH at 50°C 11
11. From 40 °C to 50 °C, the maximum % relative humidity follows the line of constant dew point.
Power line consumption
50 W max
Power voltage range 100 to 120 V, 50/60/400 Hz; 100 to 240 V
50/60 Hz
Environmental rating 0 to +50 °C
3,000 m Max
Electromagnetic compatibility Meets EMC directive (2004/108/EC), meets or exceeds IEC
61326-1:2005/EN61326-1:2013 (basic)
IEC 61000-4-2/EN
61000-4-2
IEC 61000-4-3/EN
61000-4-3
IEC 61000-4-4/EN
61000-4-4
IEC 61000-4-5/EN
61000-4-5
IEC 61000-4-6/EN
61000-4-6
IEC 61000-4-8/EN
61000-4-8
IEC 61000-4-11/EN
61000-4-11
Canada: ICES/NMB-001:2006
Australia/New Zealand: AS/NZS CISPER
11:2011
Safety ANSI/UL Std. No. 61010-1:2012; CAN/CSA-C22.2 No.
61010-1-12
ANSI/UL Std. No. 61010-2-030:2012; CAN/CSA-C22.2 No.
61010-2-030-12
Dimensions (W x H x D) 314 mm (12.4 in) x 165 mm (6.5 in) x
130 mm (5.1 in)
Weight Net: 3.23 kg (7.1 lbs.), shipping
4.2 kg (9.2 lbs.)
Display
7.0” diagonal color TFT LCD WVGA
Reference waveform display Two internal waveforms or USB thumb drive
Waveform/data storage Setups (.scp), images (.bmp, .png), channel waveforms (.csv, .bin), reference waveforms (.h5), mask (.msk),
serial protocol data (.csv), Bode gain & phase data (.csv)
Max USB flash drive size Supports industry standard flash drives
Setups without USB flash drive
10 internal setups
USB drive format FAT32 , NTFS, EXT2/3/4
Configuring your InfiniiVision
1000 X-Series Oscilloscope
Step
1: Choose your oscilloscope
EDUX1052A 50 MHz
2 channels
EDUX1052G 50 MHz
2 channels with function generator
DSOX1202A 70/100/200 MHz
2 channels
DSOX1202G 70/100/200 MHz
2 channels with function generator
DSOX1204A 70/100/200 MHz
4 channels
DSOX1204G 70/100/200 MHz
4 channels with function generator
Step
2: Select bandwidth options
Model: DSOX1202A/G (2-channel models)
DSOX1202-070
70 MHz bandwidth option Compatible with DSOX1202A or DSOX1202G
DSOX1202-100
100 MHz bandwidth option Compatible with DSOX1202A or DSOX1202G
DSOX1202-200
200 MHz bandwidth option Compatible with DSOX1202A or DSOX1202G
Model: DSOX1204A/G (4-channel models)
DSOX1200-070
70 MHz bandwidth option Compatible with DSOX1204A or DSOX1204G
DSOX1200-100
100 MHz bandwidth option Compatible with DSOX1204A or DSOX1204G
DSOX1200-200
200 MHz bandwidth option Compatible with DSOX1204A or DSOX1204G
Step
3: Select optional accessories
DSOXBODE Bode plot training kit Option
N2137A User's Guide (hardcopy) for InfiniiVision
1000 X-Series Option (electronic copy downloadable at no charge)
N2738A Soft carrying case for 1000 X-Series oscilloscopes Option
N2138A Rackmount kit for 1000 X-Series oscilloscopes Option
Step
4: Select optional PC-based test automation and documentation software
BV0004B BenchVue oscilloscope application Standard
D9010UDAA User-defined Application (UDA) software Option
D9010BSEO Infiniium Offline Oscilloscope Analysis Software Option
Configuring your InfiniiVision
1000X-Series Oscilloscope (continued)
Step
5: Select optional probes
N2142A 1:1, 10:1 switchable 75 MHz passive probe
2 probes included standard with EDUX1052A/G
N2140A 1:1, 10:1 switchable 200 MHz passive probe
2 probes included standard with DSOX1202A/G
4 probes included standard with DSOX1204A/G
N2842A 10:1
300 MHz passive probe Option
N2889A 1:1, 10:1 switchable
350 MHz passive probe Option
10070D 1:1
20 MHz passive probe Option
N2870A 1:1
35 MHz passive probe Option
N7007A 10:1
400 MHz extreme temperature passive probe Option
10076C 100:1 500 MHz
3.7 KV high voltage passive probe Option
N2791A 25 MHz, 10:1, 100:1 switchable high voltage up to
± 700V Option
N2891A 70 MHz, 100:1, 1000:1 switchable high voltage up to
± 7000V Option
Current Probes 1146B 100 kHz
100A, AC/DC current probe Option
N2780B 2 MHz
500A, AC/DC current probe (with N2779A power supply) Option
N2781B 10 MHz
150A, AC/DC current probe (with N2779A power supply) Option
N2783B 50 MHz
30A, AC/DC current probe (with N2779A power supply) Option
N2783B 100 MHz
30A, AC/DC current probe (with N2779A power supply) Option
N7040A 23 MHz
3 kA, AC current probe (Rogowski coil) Option
N7041A 30 MHz
600A, AC current probe (Rogowski coil) Option
N7042A 30 MHz
300A, AC current probe (Rogowski coil) Option
Step
6: Select language options (hard copy of user’s guide is not included unless ordered)
Front panel overlay (EDUX1052A/G, DSOX1202A/G)
Front panel overlay (DSOX1204A/G)
English Standard Standard N2137A-ABA
适用型号 Chinese (Traditional) DSOX1202-AB0 DSOX1200-AB0 N2137A-AB0
Czech DSOX1202-AKB DSOX1200-AKB Not available
适用型号 Korean DSOX1202-AB1 DSOX1200-AB1 N2137A-AB1
Polish DSOX1202-AKD DSOX1200-AKD Not available
适用型号 Spanish DSOX1202-ABE DSOX1200-ABE N2137A-ABE
条件:please contact your local Keysight office. The complete list is available at:
Thai DSOX1202-AB3 DSOX1200-AB3 Not available
Turkish DSOX1202-AB8 DSOX1200-AB8 Not available
For more information on Keysight Technologies’ products, applications or services,
This information is subject to change without notice. © Keysight Technologies, 2018 - 202 4, Published in USA, November 12, 202 4, 5992
-3484EN
Configuring your InfiniiVision
1000X-Series Oscilloscope (continued)
After-purchase bandwidth upgrades
Model: DSOX1202A/G (2-channel models)
D1202BW1A Upgrade bandwidth from 70 to 100 MHz Compatible with DSOX1202A or DSOX1202G
D1202BW2A Upgrade bandwidth from 70 to 200 MHz Compatible with DSOX1202A or DSOX1202G
D1202BW3A Upgrade bandwidth from 100 to 200 MHz Compatible with DSOX1202A or DSOX1202G
Model: DSOX1204A/G (4-channel models)
D1200BW1A Upgrade bandwidth from 70 to 100 MHz Compatible with DSOX1204A or DSOX1204G
D1200BW2A Upgrade bandwidth from 70 to 200 MHz Compatible with DSOX1204A or DSOX1204G
D1200BW3A Upgrade bandwidth from 100 to 200 MHz Compatible with DSOX1204A or DSOX1204G
Standard passive probes (Two N2142A for EDUX1052A/G;
以上指标摘自厂家官方数据手册,完整定义、测试条件与保证值口径以手册为准。完整指标、测试条件与定义口径请见产品页「软件和手册 | 白皮书」栏目中的官方数据手册。