E5260A 分析仪 Upto8 详细指标
IV measurement systems for device characterization
The GNDU is furnished with the E5260A mainframe.
Output voltage: 0 V
± 100 µV
For 1.6 A < IS ≤
2.0 A:
For 2.0 A < IS ≤
4.0 A:
Where IS is the current being sunk by the GNDU
Note: This document lists specifications and supplemental information for the E5260A and its associated modules. The specifications are the standards against which the E5260A and its associated modules are tested. When the E5260A or any of
its associated modules are shipped from the factory, they meet
the specifications. The “supplemental” information and “typical”
entries in the following specifications are not warranted, but
provide useful information about the functions and performance
二、E5260A Mainframe Specifications
Supported plug-In modules
The E5260A supports eight slots for plug-in modules.
Part number Description Slots occupied Range of operation Minimum resolution
E5290A High speed HPSMU 2 –200 V to 200 V, –1 A to 1 A 100 µV
5 pA
E5291A High speed MPSMU 1 –100 V to 100 V, –200 mA to 200 mA 100 µV
5 pA
05 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
三、MPSMU (Medium 功率 SMU) 模块规格
条件:1. ± (% of output/measured value + offset voltage)
2. 200 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), Vo is the output voltage in volts.
3. 200 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in volts.
Measure resolution Force accuracy 1 Measure accuracy
1
Maximum current ±2 V 100 µV 100 µV ±(0.03 % + 900 µV) ±(0.03 % + 700 µV) 200 mA ±20 V 1 mV 1 mV ±(0.03 % + 4 mV) ±(0.03 % + 4 mV)
200 mA
±40 V 2 mV 2 mV ±(0.03 % + 7 mV) ±(0.03 % + 8 mV)
2
±100 V 5 mV 5 mV ±(0.04 % + 15 mV) ±(0.03 % + 20 mV)
3
条件:4. ± (% of output/measured value + offset current A (fixed part determined by the output/measurement range + proportional part that is multiplied by Vo)
5. Round up below decimal point for the value of (Vo/25) and (Vo/50)
6. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in amps.
7. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 200 mA), Io is the output current in amps.
Measure resolution 4 Force accuracy 1 Measure accuracy
1,2
Maximum voltage ±100 nA 5 pA 5 pA ±(0.12 % + 50 pA + 5 pA x (Vo/25)) ±(0.1 % + 30 pA + 5 pA x (Vo/25)) 100 V ±1 µA 50 pA 50 pA ±(0.12 % + 400 pA + 50 pA x (Vo/25)) ±(0.1 % + 200 pA + 50 pA x (Vo/25))
100 V
±10 µA 500 pA 500 pA ±(0.12 % + 5 nA + 500 pA x (Vo/25)) ±(0.1 % + 3 nA + 500 pA x (Vo/25))
100 V
±100 µA 5 nA 5 nA ±(0.12 % + 40 nA + 5 nA x (Vo/25)) ±(0.1 % + 20 nA + 5 nA x (Vo/25))
100 V
±1 mA 50 nA 50 nA ±(0.12 % + 500 nA + 50 nA x (Vo/25)) ±(0.1 % + 300 nA + 50 nA x (Vo/25))
100 V
±10 mA 500 nA 500 nA ±(0.12 % + 4 µA + 500 nA x (Vo/25)) ±(0.1 % + 2 µA + 500 nA x (Vo/25))
100 V
±100 mA 5 µA 5 µA ±(0.12 % + 50 µA + 5 µA x (Vo/25)) ±(0.1 % + 30 µA + 5 µA x (Vo/25))
3
±200 mA 10 µA 10 µA ±(0.12 % + 100 µA + 10 µA x (Vo/50)) ±(0.1 % + 60 µA + 10 µA x (Vo/50))
4
-20
-50
-200
Voltage range Power 2 V 20 x Ic (W) 20 V 20 x Ic (W) 40 V
40 x Ic (W)
20 < Vc ≤ 40
40 x Io (W)
40 < Vc ≤ 100
100 x Io (W)
Where Vc is the voltage compliance setting and Io is output current.
06 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Voltage/current compliance (limiting)
The SMU can limit output voltage or current to prevent damaging
Current: ± 100 pA to
± 200 mA
Compliance accuracy: Same as the current (or voltage) set accuracy.
Maximum allowable cable resistance (Kelvin connection)
Force Line: 10 Ω (I ≤
100 mA)
Force Line: 1.5 Ω (100 mA < I ≤
200 mA)
Sense Line
10 Ω (All cases)
Voltage source output resistance
0.3 Ω typical (Force line
Voltage measurement input resistance: ≥ 10
13 Ω
Current source output resistance: ≥ 10
13 Ω (1 nA range)
Current compliance setting accuracy (for opposite polarity)
For 100 nA to 200 mA ranges: I setting accuracy
± 2.5 % of range
For 100 nA to 10 mA ranges
10 nF
For 100 mA to 200 mA ranges
100 µF
Maximum guard capacitance
900 pF
Maximum shield capacitance
5000 pF
Maximum guard offset voltage
± 1 mV
Voltage source
0.01 % of V range (rms)
Current source
0.1 % of I range (rms)
Voltage source
0.03 % of V range
Current source
1 % of I range
范围 switching transient 噪声 (典型值, 滤波器 ON)
Trigger out delay (pulsed measurements)
0 to 32.7 ms with 100 µs resolution (< pulse width)
条件:1. ± (% of output/measured value + offset voltage V)
2. 1 A (Vo ≤ 20 V), 500 mA (20 V <Vo ≤ 40 V), Vo is the output voltage in volts.
3. 1 A (Vo ≤ 20 V), 500 mA (20 V <Vo ≤ 40 V), 125 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in volts.
4. 1 A (Vo ≤ 20 V), 500 mA (20 V <Vo ≤ 40 V), 125 mA (40 V < Vo ≤ 100 V), 50 mA (100 V < Vo ≤ 200 V), Vo is the output voltage in volts.
Force resolution Measure resolution Force accuracy 1 Measure accuracy
1
Maximum current ±2 V 100 µV 100 µV ±(0.03 % + 900 µV) ±(0.03 % + 700 µV) 1 A ±20 V 1 mV 1 mV ±(0.03 % + 4 mV) ±(0.03 % + 4 mV)
1 A
±40 V 2 mV 2 mV ±(0.03 % + 7 mV) ±(0.03 % + 8 mV)
2
±100 V 5 mV 5 mV ±(0.04 % + 15 mV) ±(0.03 % + 20 mV)
3
±200 V 10 mV 10 mV ±(0.045 % + 30 mV) ±(0.035 % + 40 mV)
4
条件:5. ± (% of output/measured value + offset current A (fixed part determined by the output/measurement range + proportional part that is multiplied by Vo)
6. Round up below decimal point for the value of (Vo/25).
7. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 100 mA)
8. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 125 mA), 40 V (125 mA < Io ≤ 500 mA), 20 V (500 mA < Io ≤ 1 A), Io is the output current in amps.
Measure resolution Force accuracy 1,2 Measure accuracy
1,2
Maximum voltage ±100 nA 5 pA 5 pA ±(0.12 % + 50 pA + 5 pA x (Vo/25)) ±(0.1 % + 30 pA + 5 pA x (Vo/25)) 200 V ±1 µA 50 pA 50 pA ±(0.12 % + 400 pA + 50 pA x (Vo/25)) ±(0.1 % + 200 pA + 50 pA x (Vo/25))
200 V
±10 µA 500 pA 500 pA ±(0.12 % + 5 nA + 500 pA x (Vo/25)) ±(0.1 % + 3 nA + 500 pA x (Vo/25))
200 V
±100 µA 5 nA 5 nA ±(0.12 % + 40 nA + 5 nA x (Vo/25)) ±(0.1 % + 20 nA + 5 nA x (Vo/25))
200 V
±1 mA 50 nA 50 nA ±(0.12 % + 500 nA + 50 nA x (Vo/25)) ±(0.1 % + 300 nA + 50 nA x (Vo/25))
200 V
±10 mA 500 nA 500 nA ±(0.12 % + 4 µA + 500 nA x (Vo/25)) ±(0.1 % + 2 µA + 500 nA x (Vo/25))
200 V
±100 mA 5 µA 5 µA ±(0.12 % + 50 µA + 5 µA x (Vo/25)) ±(0.1 % + 30 µA + 5 µA x (Vo/25))
3
±1 A 50 µA 50 µA ±(0.5 % + 500 µA + 50 µA x (Vo/25)) ±(0.5 % + 300 µA + 50 µA x (Vo/25))
4
07 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Voltage range Power 2 V 20 x Ic (W) 20 V 20 x Ic (W) 40 V
40 x Ic (W)
20 < Vc ≤ 40
40 x Io (W)
40 < Vc ≤ 100
100 x Io (W)
100 < Vc ≤ 200
200 x Io (W)
Where Vc is the voltage compliance setting and Io is output current.
HPSMU measurement and output range
1000
-50
-125
-500
-1000
-200 -100 -40 -20 20 40 100
200
Voltage/current compliance(limiting)
The SMU can limit output voltage or current to prevent damaging
Current: ± 100 pA to
± 1 A
Compliance accuracy: Same as the current (or voltage) set accuracy.
Maximum allowable cable resistance (Kelvin connection)
Force line: 10 Ω (I ≤
100 mA)
Force line: 1.5 Ω (100 mA < I ≤
1 A)
Sense line
10 Ω (All cases)
Voltage source output resistance
0.2 Ω typical (Force line, non-Kelvin connection)
Voltage measurement input resistance: ≥
1013 Ω
Current source output resistance: ≥
1013 Ω (1 nA range)
Current compliance setting accuracy (for opposite polarity)
For 1 nA to 10 nA ranges: I setting accuracy
± 12 % of range
For 100 nA to 1 A ranges: I setting accuracy
± 2.5 % of range
For 100 nA to 10 mA ranges
10 nF
For 100 mA to 1 A ranges
100 µF
Maximum guard capacitance
900 pF
Maximum shield capacitance
5000 pF
Maximum guard offset voltage
± 1 mV
Voltage source
0.01 % of V range (rms)
Current source
0.1 % of I range (rms)
Voltage source
0.03 % of V range
Current source
1 % of I range
范围 switching transient 噪声 (典型值, 滤波器 ON)
Trigger out delay (pulsed measurements): 0 to 32.7 ms with 100 µs resolution (< pulse width)
08 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Outputs swept voltage or current, and measures dc voltage or current. One channel can sweep current or voltage while up to eight channels can measure current or voltage. A second chan - nel can be synchronized with the primary sweep channel as an
additional voltage or current sweep source. Linear or log sweeps
Hold time: 0 – 655.35 s
1 ms resolution
Delay time: 0 – 65.5350 s
100 µs resolution
条件:Outputs swept voltage or current, and measures dc voltage or current. Up to eight channels can sweep current or voltage and up to eight channels can measure current or voltage. Linear or log sweeps can be performed.
Hold time: 0 – 655.35 s
1 ms resolution
Delay time: 0 – 65.5350 s
100 µs resolution
Outputs a voltage or current pulse and measures dc voltage or
Pulse width: 500 µs to 100 ms
100 µs resolution
Pulse period: 5 ms to 1 s (≥ pulse width + 4 ms)
100 µs resolution
Outputs pulsed swept voltage or current, and measures dc voltage or
current. A second channel can be programmed to output a staircase
sweep voltage or current synchronized with the pulsed sweep output.
Staircase 扫描(含) 脉冲 bias 测量模式
Outputs swept voltage or current, and measures dc voltage or current. A second channel can be programmed to output a pulsed bias voltage or current. A third channel can be synchro - nized with the primary sweep channel as an additional voltage
Outputs quasi-pulsed voltage and measures dc voltage or current.
Outputs and measure voltage or current by using linear search
Outputs and measure voltage or current by using binary search
The E5260A supports a time stamp function utilizing an internal
The E5260A mainframe contains (volatile) memory that can be
used to increase test measurement throughput. Program memory allows the storage of program code in the E5260A, eliminating the need to communicate over the GPIB interface. In addition, input data can be passed to code sequences stored in program
Maximum lines of storable code
40,000
Maximum number of program sequences
2,000
The number of data points that can be stored in the data buffer
varies with the choice of the output data format
Minimum number of storable data Points
34,034
Trigger in/out synchronization pulses before and after setting and
measuring dc voltage and current. Arbitrary trigger events can be
masked or activated independently.
09 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
条件:3. Start the source output at each sweep step for a staircase
sweep, pulsed sweep, staircase sweep with pulsed bias, or multi-channel sweep measurement. 4. Start the pulsed output for a pulsed spot measurement. 5. Recover from a wait state.
An external trigger input signal can be used to do any of the
following: 1. Start a measurement
2. Start a measurement at each sweep step for a staircase sweep or multi channel sweep measurement
Input level: TTL level, negative or positive edge trigger, or TTL
条件:events occurs: 1. The end of a measurement is reached. 2. The end of a measurement at each sweep step for a staircase sweep or multi-channel sweep measurement is reached.
3. Completion of the source output setup at each sweep step
for a staircase sweep, pulsed sweep, staircase sweep with pulsed bias, or multi-channel sweep measurement. 4. Completion of the pulsed output setup for a pulsed spot measurement.
5. A trigger command is issued.
An output trigger signal can be sent when one of the following
Output level: TTL level, negative or positive edge trigger, or
TTL level, negative or positive gate trigger
条件:with other instruments. They can also be used as output and input ports for digital signals. The user can selectively assign pins to trigger mode or digital I/O mode.
General Purpose Digital I/O 16 general-purpose digital I/O signals are available via a
25-pin
DIN connector. These pins can be used as an alternative to the
BNC trigger-in and trigger-out lines to synchronize the E5260A
Operating
15 % to 80 % RH, non-condensing
Storage
5 % to 90 % RH, non-condensing
Operating
0 m to 2,000 m (6,561 ft)
Storage
0 m to 4,600 m (15,092 ft)
Line frequency
47 Hz to 63 Hz
适用型号 EMC: IEC61326-1/EN61326-1
KC: RRA Notification amending Radio Waves
适用型号安全: IEC61010-1/EN61010-1
CAN/CSA-C22.2 No.
61010-1-12, C/US
E5260A: 426 mm W x 235 mm H x
575 mm D
E5260A mainframe only
17 kg
– USB-GPIB interface (Keysight
82357B)
– GNDU to Kelvin adaptor (Keysight N1254A-100)
– CD-ROMs (EasyEXPERT install media, and VXI plug&play driver)
Supported OS: Windows
7 Professional
10 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Keysight EasyEXPERT group+ Software
条件:allows you the option of storing test condition and measurement data automatically
valuable information is not lost and that measurements can be repeated at a later date.
environment that facilitate the development of complex testing algorithms.
Keysight EasyEXPERT group+ GUI based characterization software is available on
your PC to accelerate the characterization tasks. It supports efficient and repeatable
device characterization in the entire characterization process from measurement
setup and execution to analysis and data management either by interactive manual
operation or automation across a wafer in conjunction with a semiautomatic wafer
prober. EasyEXPERT group+ makes it easy to perform complex device characterization
immediately with the ready-to-use measurements (application tests) furnished, and
after each measurement in a unique built-in database (workspace), ensuring that
Finally, EasyEXPERT has built-in analysis capabilities and a graphical programming
– Multiple measurement modes for quick setup and measurement execution
(application 测试, classic 测试, 和 quick 测试)
– Graphical display, automated analysis capabilities and data generation to
Excel and image for analysis and reporting
– Built-in database (workspace) records test data automatically and simplifies
the data management without numerous data files
– GUI-based control of the Keysight B2200A, B2201A and E5250A
– EasyEXPERT remote control function supports the remote measurement
execution of application tests that are created on GUI interactively, via the
– Data back capability and various data protection feature for shared usage
– Characterization environment is available on user’s PC as a personal and
portable analyzer environment. EasyEXPERT group+ can be installed on any
EasyEXPERT group+ comes with the application tests conveniently organized by device
type, application, and technology. You can easily edit and customize the furnished appli -
cation tests to fit your specific needs. Application tests are provided for the following
CMOS Transistor Id-Vg, Id-Vd, Vth, breakdown, etc
Bipolar Transistor Ic-Vc, diode, Gummel plot, breakdown, hfe, etc.
Nano Device Resistance, Id-Vg, Id-Vd, Ic-Vc, etc
条件:Reliability test NBTI/PBTI, electro migration, hot carrier injection, J-Ramp, TDDB, etc.
11 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
The application test mode provides application oriented point-
and-click test setup and execution. An application test can be se -
lected from the library by device type and desired measurement,
and then executed after modifying the default input parameters
as needed. Available application tests can be adapted to config -
The tracer test mode offers intuitive and interactive sweep control
using a rotary knob similar to a curve tracer. Just like an analog
curve tracer, you can sweep in only one direction (useful for R&D
device analysis) or in both directions (useful in failure analysis
applications). Test set ups created in tracer test mode can be
seamlessly and instantaneously transferred to classic test mode
A GUI-based Quick Test mode enables you to perform test se - quencing without programming. You can select, copy, rearrange and cut-and-paste any application tests with a few simple mouse clicks. Once you have selected and arranged your tests, simply
click on the measurement button to begin running an automated
Limited auto ranging, voltage/current compliance, power compliance, automatic sweep abort functions, self-test, and self-calibration
SMUs in “Standby” remain programmed to their specified output
value even as other units are reset for the next measurement.
条件:measurement. The function ceases as soon as these conditions end or when a measurement that does not use this function is started.
This function allows you to keep a source active between measurements. The source module will apply the specified bias between measurements when running classic tests inside an application test, in quick test mode, or during a repeated
This function subtracts the offset current from the current measurement raw data, and returns the result as the measure - ment data. This function is used to compensate the error factor (offset current) caused by the measurement path such as the
X-axis and up to eight Y-axes, linear and log scale, real time
Marker: Marker to min/max, interpolation, direct marker, and marker skip
Line: Two lines, normal mode, grad mode, tangent mode, and regression mode
Measurement data and calculated user function data are listed
in conjunction with sweep step number or time domain sampling
step number. Up to 20 data sets can be displayed.
Up to 20 user-defined parameters can be displayed on the
Automatic analysis function
On a graphics plot, the markers and lines can be automatically
located using the auto analysis setup. Parameters can be auto -
matically determined using automatic analysis, user function,
Up to 20 user-defined analysis functions can be defined using
The read out functions are built-in functions for reading various
X-Y graph plot can be printed or stored as image data to clipboard or mass storage device. (File type: bmp, gif, png, emf). Graph and list data can be exported to Excel.
12 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Up to 20 user-defined functions can be defined using
条件:base e), mavg (moving average), max,min, sqrt, trigonometric function, inverse trigonometric function, and so on.
+, -, *, /, ^, abs (absolute value), at (arc tangent), avg (averaging),
cond (conditional evaluation), delta, diff (differential), exp (expo -
nent), integ (integration), lgt (logarithm, base
10), log (logarithm
Keyboard constants are stored in memory as follows: q: Electron charge, 1.602177E-19 C k: Boltzman’s constant
1.380658E-23 ε (e): Dielectric constant of vacuum
8.854188E-12
The following unit symbols are also available on the keyboard:
a (10-18), f (10-15), p (10-12), n (10-9),
u or μ (10-6), m (10-3), k (103), M (106),
G (109), T (1012) , P (1015)
EasyEXPERT group+ supports the built-in database called
条件:“workspace”. Workspaces are created on a HDD, and they enable to manage and access all the measurement related data without handling numerous files. Every workspace supports the following features:
– Access to measurement capabilities and data stored in the workspace.
– Save/Import/Export measurement settings and data (application library, measurement settings, my favorite setup, and measurement data)
– Recall the setup for measurement reproduction and data for analysis
Workspace back-up and portability
EasyEXPERT group+ has the ability to import/export a workspace for back-up and portability.
13 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet
Precision 电流 - 电压分析仪 Series
Discontinued Advanced Device Analyzer Precision IV Analyzer Economic IV
I/V Sweep Yes Yes Yes Yes Yes Yes1
GUI based switching matrix control
Yes2 - Yes2 Yes2 Yes2 Yes2
Oscilloscope view Yes3 Yes3 - - - -
External instrument driver support
LCR meter (4284A/E4980A)
Pulse Generator (81110A) Yes Yes Yes Yes Yes Yes
DVM (3458A) Yes Yes Yes Yes Yes Yes
Prober control in Quick Test mode Yes4 Yes4 Yes4 Yes4 Yes4 Yes4
Firmware requirement A.04.00 or
later5
A.04.00 or
later5
B.01.10 or
条件:later 1. PGU and VSU/VMU are supported. Differential voltage measurement of VMU is not supported. 2. B2200/01A and E5250A (with E5252A cards) are supported 3. Only available for supported modules.
4. Cascade Microtech Sumit 12000/S300 (Nucleus), Cascade Microtech (Suss MicroTec) PA200/PA300, and Vector Semiconductor VX-2000/VX-3000
5. The latest firmware version is strongly recommended to take full advantage of measurement capabilities.
B.01.10 or
later 1.0 or later HOSTC
03.08 or later
14 | Keysight | E5260A IV Analyzer/8 Slot High Speed Measurement Mainframe - Data Sheet