松田精密|Matsusada Precision Inc.
AU-100P1.5 直流电源 详细指标
Low Profile/High Power Rack-Mount Type High Voltage Power Supply 100kV / 1.5mA / 150W
详细指标
以下指标取自松田精密(Matsusada Precision)官网 AU 系列产品页的「Models / Specifications / Options / Dimensions」区块,其中「型号输出参数」为该型号在官方型号表中的原始列值;官方原文为英文,此处按原文保留。详细规格与测试条件以官方数据手册为准。
型号输出参数
• 输出电压:+100 kV
• 电流:1.5 mA
• 功率:150 W
系列共用规格
• 输入电压:100 to 120 Vac, 50/60 Hz, Single-phase
200 to 240 Vac, 50/60 Hz, Single-phase
200 to 240 Vac, 50/60 Hz, Three-phase (option)
• 输出电压控制:[Local] 10-turn potentiometer on front panel
[Analog remote] External control voltage 0 to 10 Vdc
[Digital remote] Additional CO-HV adapter
• 输出电流控制:[Local] 10-turn potentiometer on front panel
[Analog remote] External control voltage 0 to 10 Vdc
[Digital remote] Additional CO-HV adapter
选件 / 附件
• -LC:Enables constant voltage (CV) and constant current (CC) operation modes. *1
Current regulation 0.05% (-LC option eliminates overload trip function)
• -LF:Floating ground (withstanding voltage of 50 Vdc) *2
Used when measuring minimal current in the load. All equipment that connects to the Remote control connector (TB1) must be on the floating ground in case this feature is intended to be used. (Cannot be used for the purpose of floating high-voltage power supply.)
• -LMs:Master-Slave control (0.6 kW, 1.2 kW, and 2.2 kW models only) *1 *2 *3
A maximum of four slave units can be controlled with one master unit.
• -LPs:Setting value display
The set voltage and current values are displayed on the meter while you press the SET button on the front panel. (This button is attached only when the -LPs option is chosen.)
• -LStc:Constant current status signal
In constant current mode or when Overcurrent Protection (OCP) is activated, the open collector will transition to a high state. (V CE > 2 V)
• -LStv:Constant voltage status signal
When the operation mode is Constant voltage mode, the open collector will transition to a high state. (V CE > 2 V)
• -LW:Slow start *2
Takes around 10 seconds from turning on the OUTPUT switch, external output ON/OFF, and remote reset to reach the maximum rated voltage
(about 5 seconds at half the maximum rated voltage as output voltage).
• -L(U):Input voltage switch over 100 Vac to 120 Vac/200 Vac to 240 Vac single phase input. Internal switch over.
(30 W to 300 W models only)
Two AC input cables (CABLE TYPE 1 and CABLE TYPE 3) are included.
• -L(115V):100 Vac to 120 Vac single phase input.
(600 W models only) (CE not compatible)
• -L(220V):200 Vac to 240 Vac single phase input.
This option is available only for models with 30W to 300W output power.
The CABLE TYPE 3 AC input cable is included.
• -L(200V3P):three-phase input
This option is available only for 2200W output models.
No AC input cable is included. Please purchase CABLE TYPE 6 separately.
• -L(3m):Change the length of the High-voltage output shielded cable to 3 meters.
• -L(5m):Change the length of the High-voltage output shielded cable to 5 meters.
(only for ≤40 kV models)
• -L(7m):Change the length of the High-voltage output shielded cable to 7 meters.
(only for ≤ 15 kV models)
If the power supply is configured to cut off the output during an overcurrent event with master-slave connection, do not select -LC option for Master unit (the other options can be selected), and select -LC option for only slave unit (the other options can also be selected together). Combinations other than the above cut off the output when overcurrent will not work. Also, the slave unit is equipped with the -LC option. Therefore, if the slave unit is used individually, output will be either CV or CC.
In case selecting -LMs option with -LF option or -LW option, all AU power supplies connected in a master-slave configuration must be equipped with the -LF option or -LW option.
In case you operate power supplies with CV/CC mode under master/slave connection, select the -LC option for the master unit and all slave units (the other options can also be selected together). Combinations other than the above, CV/CC will not work. Also, the Master unit and each Slave unit are equipped with the -LC option. Therefore, if each unit is used individually, the output will be either CV or CC.
• 光通信:Optical isolation adapters utilize fiber optic cables for digital communication, which enables extremely reliable communication even in noisy environments such as in factories and for long-distance control.
By using optical fibers, they are electrically isolated, so safe operation is possible even in power supply configurations with potential differences.
Application
Control of high-voltage power supplies via LAN, USB, RS-232C, RS-485, or GPIB.
Connection
Using fiber-optic cables, the adapter for each interface is connected to CO-HV units.
And with the dedicated control cable, the CO-HV units are connected to a high-voltage power supply AU series.
Number of units
Up to 32 pairs with CO-HV units and high-voltage power supplies are available.
With GPIB, by setting the address, you can connect a maximum of 448 high-voltage power supply units (14 addresses x 32 units).
Features
The communication system is well suited for such situations where there is a distance between the computer and the high-voltage power supply or in noisy environments. Also, it is especially ideal for use in combination with DC power supplies. The standard cable length is 2 meters, and it can be optionally extended up to 40 meters.
LAN
Order both CO-E32 and CO-HV (with CO-AU cable) together.
USB
Order both USB-OPT and CO-HV (with CO-AU cable) together.
RS-232C and RS-485
Order both CO-OPT2-9 , or CO-OPT4-25 along with CO-HV (with CO-AU cable) together.
GPIB
Order both CO-G32 and CO-HV (with CO-AU cable) together.
• 输入电缆:Standard
30 W to 300 W models or
-L(U) option
CABLE TYPE1
Rated voltage: 125 V
Maximum current: 10 A
For Japan and North America
2.5 meters
Fixed length
Standard
30 W to 300 W models with
-L(U) option, -L(220V) option
CABLE TYPE3
Rated voltage: 250 V
Maximum current: 10 A
For European nations (VDE, TUV)
2.5 meters
Fixed length
Sold separately
30 W to 300 W models with
-L(220V) option
CABLE TYPE4
Rated voltage: 300 V
Maximum current: 25 A
For European nations (VDE)
2.5 meters Fixed length
Standard
600 W, 1200 W, 2200 W models
CABLE TYPE5
Rated voltage: 300V
Maximum current: 25 A
For Japan and North America
2.5 meters
Sold separately
2200 W models with
-L(200V3P) option
CABLE TYPE6
Rated voltage: 600V
Maximum current: 25 A
For Japan only
2.5 meters
As for the AC input cable, select and use an AC input cable suitable for the usage environment and area. If none of the following applies, please source a suitable cable that meets your local requirements and conditions.
• 输出电缆:Standard 1 kV to 15 kV models (except for 2200 W models
[1.5 kV or less])
CN-15-MHVP
2.5 meters
Specified length available
Standard 2200 W models (1kV, 1.5kV)
CN-15PR-MHVP
2.5 meters Fixed length
Standard 20 kV to 30 kV models
CN-30-MHVP
2.5meters
Specified length available
Standard 40 kV models
CN-45-MHVP
2.5meters Specified length available
Standard 50 kV to 60 kV models
CN-60-MHVP
2.5meters Specified length available
Standard 80 kV or more models
CN-120AR-MHVP
2.5 meters Specified length available
• 应用软件:PSS2en series: Remote control, Test workflow design, and Data logging
PSS2en is the dedicated software that can operate various power supplies, electronic loads, and digital controllers for power supplies manufactured by Matsusada Precision Inc. with a simple setup.
It is perfect for the aging test, the burn-in test, and the withstand voltage test for electronic parts, as well as for the endurance test, intermittent/continuous operation test, or various simulation tests for automobile electric components.
Click here for the PSS2en series
资源下载
以上规格取自松田精密(Matsusada Precision)官网 AU 系列产品页(https://www.matsusada.com/product/high-voltage-power-supplies/rack-mount/au/)。
松田精密(Matsusada Precision) · onetest 仪器资源库